X Ray Hand Switch
Model: L01A
Highlight
- The x ray hand switch of L01A with two steps, it is mainly designed for different kinds of x ray to control the x ray exposure.
- Two steps design, when press step one, the x ray machine enters into x ray exposure ready status, and when the two steps press together, the x ray machine finishes exposure.
- Special inner micro switch design which ensure the using life and electrical life longer and stable.
- The spring coil wire includes 3-core or 4- core, the customer could choose freely.
- As for the x ray push button switch holder, it has two types for option, one with hole and the other without.
- It is applicable to portable x ray, mobile x ray, digital x ray, fixed x ray, stationary x ray, c-arm, uc-arm, analog x ray, veterinary x ray, industrial x ray etc.
- It is also used on beauty machine and equipment.
- The hand switch connector, such as crystal, aviation plug etc which can be customized.
- It is compatible to omron push button switches C2U. And it can be replaced for GE, Canon, Siemens, Hitachi, Fuji, Agfa, Philips, Carestream, Minxray, Shamadzu, Envista etc x ray brand equipment.
Pictures
Specification
1. Mechanical parameters
- mechanical life: ≤ 200, 000.00 times
- handle size: length 10.5 cm (max), diameter 3 cm
- spring coil wire: 4-core or 3-core opitonal
- spring coil wire length: standard configuration of 3m, 6m and 10m. Other length can be customized.
- 2 steps control.
2. Electrical parameters
- electrical using life: ≤ 400, 000.00 times
- operation voltage: AC 125V 1A, DC 30V 2A
3. Style
- smooth surface
4. Wire connection method
- step 1: connect the black and white wires
- step 2: connect the red and green wires
Online Service